Arietta 65 Ultrasound Software Versions: V1.0 through V4.0.0
Check against the official notice below — it lists the full affected range.
FujiFilm Healthcare Americas Corporation is recalling Arietta 65 Ultrasound systems with software versions V1.0 through V4.0.0 that were distributed nationwide in the US. When used with Fujifilm (formerly Hitachi) transducers, these systems contain an error in the calculation formula for Effective Regurgitant Orifice Area (EROA) when performing MR Flow combined with Left Ventricular Outflow Tract Flow (LVOT Flow) after an echocardiogram. This error causes incorrect values to be displayed to the user, and the same error also appears in the instruction manual. A total of 48 systems with specific serial numbers are affected by this recall.
The recall notice does not specify what remedy consumers should take or what steps they should follow.
Written by software from the official notice below. Not reviewed by a lawyer.
When the ultrasound system is used in combination with a Fujifilm (formally Hitachi) transducer, there is an error in the calculation formula of Effective Regurgitant Orifice Area (EROA) when you perform MR Flow combined with Left Ventricular Outflow Tract Flow (LVOT Flow) after performing an echocardiogram. The incorrect value is displayed. In addition, the error appears in the instruction manual
Reason for recall: When the ultrasound system is used in combination with a Fujifilm (formally Hitachi) transducer, there is an error in the calculation formula of Effective Regurgitant Orifice Area (EROA) when you perform MR Flow combined with Left Ventricular Outflow Tract Flow (LVOT Flow) after performing an echocardiogram. The incorrect value is displayed. In addition, the error appears in the instruction manual Distribution: US Nationwide distribution. Quantity: 48 US Lot/code info: Serial Numbers/UDI: G3024453 (01)04562122959278(11)191029(21)G3024453 G3024456 (01)04562122959278(11)191029(21)G3024456 G3042466 (01)04562122959278(11)191115(21)G3042466 G3042471 (01)04562122959278(11)191115(21)G3042471 G3042477 (01)04562122959278(11)191115(21)G3042477 G3042478 (01)04562122959278(11)191115(21)G3042478 G3042484 (01)04562122959278(11)191115(21)G3042484 G3042486 (01)04562122959278(11)191115(21)G3042486 G3042487 (01)04562122959278(11)191115(21)G3042487 G3042496 (01)04562122959278(11)191115(21)G3042496 G3042499 (01)04562122959278(11)191115(21)G3042499 G3042501 (01)04562122959278(11)191115(21)G3042501 G3042576 (01)04562122959278(11)191220(21)G3042576 G3042578 (01)04562122959278(11)191220(21)G3042578 G3042583 (01)04562122959278(11)191220(21)G3042583 G3042584 (01)04562122959278(11)191205(21)G3042584 G3042585 (01)04562122959278(11)191205(21)G3042585 G3042587 (01)04562122959278(11)191205(21)G3042587 G3042588 (01)04562122959278(11)191220(21)G3042588 G3082956 (01)04562122959278(11)200929(21)G3082956 G3085640 (01)04562122959278(11)200929(21)G3085640 G3085644 (01)04562122959278(11)200929(21)G3085644 G3095193 (01)04562122959278(11)201211(21)G3095193 G3095194 (01)04562122959278(11)201211(21)G3095194 G3095195 (01)04562122959278(11)201211(21)G3095195 G3095197 (01)04562122959278(11)201211(21)G3095197 G3095202 (01)04562122959278(11)201211(21)G3095202 G3095445 (01)04562122959278(11)201228(21)G3095445 G3111596 (01)04562122959278(11)210401(21)G3111596 G3111597 (01)04562122959278(11
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.