← Search recalls
FDA (Device enforcement)Recalled 2025-09-04class ii

American Contract Systems Inc.: Medical convenience kits IHDC30R D AND C PACK

Hazard

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Details from the source

Reason for recall: Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed. Distribution: US Nationwide distribution in the states of SD, IA, MN, WA, IL. Quantity: 32,433 total Lot/code info: IHDC30R UDI-DI 191072226310 lot 8096511 bag serial number N/A * If bag serial number is listed as "N/A", then whole lot is affected

UPC codes

191072226310

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify