Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
Reason for recall: Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed. Distribution: US Nationwide distribution in the states of SD, IA, MN, WA, IL. Quantity: 32,433 total Lot/code info: IHDC30R UDI-DI 191072226310 lot 8096511 bag serial number N/A * If bag serial number is listed as "N/A", then whole lot is affected
191072226310
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.