Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
Reason for recall: Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed. Distribution: US Nationwide distribution in the states of SD, IA, MN, WA, IL. Quantity: 32,433 total Lot/code info: ANLM40AG UDI-DI 191072232076 Lot 8192111 Bag serial number 68195596; ANPP15AR UDI-DI 191072214423 Lot 8024911 Bag serial number 68055493; HSLM53AB UDI-DI 191072224095 Lot 8061111 Bag serial number 68138767; MILM82A UDI-DI 191072231512 Lot 8361111 Bag serial number 68576542; RCNA47B UDI-DI 191072228536 Lot 8441711 Bag serial number 68680026; SFLM54AB UDI-DI 191072217400 lot 8105511 Bag serial number N/A; UDLM94AA UDI-DI 191072214089 Lot 8382011 Bag serial number 17621919; UILM77AU UDI-DI 191072210333 Lot 8056311 Bag serial number N/A; UINR44P UDI-DI 191072210340 Lot 8067511 Bag serial numbers 68138618 68140374; UISH15AV UDI-DI 191072227843 Lot 8028211 Bag serial number 68154067 * If bag serial number is listed as "N/A", then whole lot is affected
191072232076, 191072214423, 191072224095, 191072231512, 191072228536, 191072217400, 191072214089, 191072210333, 191072210340, 191072227843
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.