← Search recalls
FDA (Device enforcement)Recalled 2025-09-04class ii

American Contract Systems Inc.: Medical convenience kits Item Number/Description HSAN63N ANGIO PACK IHAN02W ANGIOGRAPHY PACK MHCA98AK CORONARY ANGIO PACK- 205991 RCAN45C ANGIO CUSTOM RCH UDCO10F CVL C

Hazard

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Details from the source

Reason for recall: Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed. Distribution: US Nationwide distribution in the states of SD, IA, MN, WA, IL. Quantity: 32,433 total Lot/code info: ANCA80AP UDI-DI 191072232168 Lot 8077111 Bag serial number 68119740; HSAN63N UDI-DI 191072232267 Lot 8097211 Bag serial number N/A; IHAN02W UDI-DI 191072201928 Lot 7961811 Bag serial number 68086881 Lot 8037411 Bag serial number N/A; MHCA98AK UDI-DI 191072216014 Lot 8006711 Bag serial number 47415134 Lot 8133011 Bag serial number N/A; RCAN45C UDI-DI 191072229632 Lot 8537111 Bag serial number 68872542; UDCO10F UDI-DI 191072216236 Lot 7987811 Bag serial number N/A * If bag serial number is listed as "N/A", then whole lot is affected

UPC codes

191072232168, 191072232267, 191072201928, 191072216014, 191072229632, 191072216236

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.