Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
Reason for recall: Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed. Distribution: US Nationwide distribution in the states of SD, IA, MN, WA, IL. Quantity: 32,433 total Lot/code info: RCEN25B UDI-DI 191072225115 lot 8138311 bag serial number N/A; RCHN27D UDI-DI 191072235923 lot 8508211 bag serial number 47847459 * If bag serial number is listed as "n/a", then entire lot is affected
191072225115, 191072235923
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.