Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
Reason for recall: Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed. Distribution: US Nationwide distribution in the states of SD, IA, MN, WA, IL. Quantity: 32,433 total Lot/code info: AHLD15X UDI-DI 191072224767 Lot 8188711 Bag serial no. 17555039 17555044 17555049 17609166 17609181; AKDV96B UDI-DI 191072224019 Lot 7311311 Bag serial no. 47140875 47141919; EVBH28D UDI-DI 191072239051 Lot 8424611 Bag serial no. 17754395; UPVD01K UDI-DI 191072238993 Lot 8439111 Bag serial no. 17682598 Log 8499211 Bag serial no. 17742543
191072224767, 191072224019, 191072239051, 191072238993
Official record: FDA (Device enforcement) ↗