← Search recalls
FDA (Device enforcement)Recalled 2022-10-03class ii

Becton, Dickinson and Company, BD Biosciences: BD Trucount Tubes (Cat. No. 663028), used for determining absolute counts of leukocytes in blood. BD Trucount Tubes are designed for use with in vitro di

Hazard

Due to partial label detachment on tubes. This can cause tubes to become stuck together in the pouch or cause automation errors when used with the automated loader for the wash assistant. Firm root cause investigation confirmed that label detachment is due to an adhesive change by a tier 2 supplier.

Details from the source

Reason for recall: Due to partial label detachment on tubes. This can cause tubes to become stuck together in the pouch or cause automation errors when used with the automated loader for the wash assistant. Firm root cause investigation confirmed that label detachment is due to an adhesive change by a tier 2 supplier. Distribution: U.S.: AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SD, TN, TX, UT, VA, VT, WA, and WI O.U.S.: Argentina, Australia, Belgium, Bolivia, Brazil, Canada, Chile, China, Columbia, Costa Rica, Dominican Republic, Ecuador, El Salvador, Guatemala, Hong Kong, India, Japan, Korea, Mexico, Malaysia, New Zealand, Pan Quantity: 11,709 tubes (corrected 11/04/2022) Lot/code info: Corrected as of 11/04/2022: Catalog Number: 663028 UDI-DI Code: 00382906630285 Lot Numbers: 2031238 2063994 2095615 2145791 2221697 2237045

UPC codes

00382906630285

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify