← Search recalls
FDA (Device enforcement)Recalled 2022-09-07class ii

Johnson & Johnson Surgical Vision, Inc.: THERMOCOOL SMARTTOUCH SF Bi-Directional Navigation Catheter, REF: D-1348-05-S. Used for catheter-based cardiac electrophysiological mapping (stimulation and re

Hazard

Manufacturing error resulted in potential contamination of Bi-Directional navigation catheter irrigation path with cellulose fibers that can cause 1) Failure of initial flushing inspection, 2) Insufficient irrigation causing catheter tip temperature increase, limiting the ability to deliver ablation, or 3) particles flushed through the catheter irrigation may have thrombogenic/toxic effect.

Details from the source

Reason for recall: Manufacturing error resulted in potential contamination of Bi-Directional navigation catheter irrigation path with cellulose fibers that can cause 1) Failure of initial flushing inspection, 2) Insufficient irrigation causing catheter tip temperature increase, limiting the ability to deliver ablation, or 3) particles flushed through the catheter irrigation may have thrombogenic/toxic effect. Distribution: Worldwide - US Nationwide distribution in the states of NC, TN, VA, OH, MO, MI, GA, OR, MO, SD, CO, OH, FL, NJ, TX, IL, WI and the countries of CROATIA, GERMANY, HUNGARY. Quantity: 91 Lot/code info: UDI-DI/Lots: 10846835010183/30779056L, 30808001L, 30814782L, 30814783L, 30779172L

UPC codes

10846835010183

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify