Manufacturing error resulted in potential contamination of Bi-Directional navigation catheter irrigation path with cellulose fibers that can cause 1) Failure of initial flushing inspection, 2) Insufficient irrigation causing catheter tip temperature increase, limiting the ability to deliver ablation, or 3) particles flushed through the catheter irrigation may have thrombogenic/toxic effect.
Reason for recall: Manufacturing error resulted in potential contamination of Bi-Directional navigation catheter irrigation path with cellulose fibers that can cause 1) Failure of initial flushing inspection, 2) Insufficient irrigation causing catheter tip temperature increase, limiting the ability to deliver ablation, or 3) particles flushed through the catheter irrigation may have thrombogenic/toxic effect. Distribution: Worldwide - US Nationwide distribution in the states of NC, TN, VA, OH, MO, MI, GA, OR, MO, SD, CO, OH, FL, NJ, TX, IL, WI and the countries of CROATIA, GERMANY, HUNGARY. Quantity: 91 Lot/code info: UDI-DI/Lots: 10846835010183/30779056L, 30808001L, 30814782L, 30814783L, 30779172L
10846835010183
Official record: FDA (Device enforcement) ↗