← Search recalls
FDA (Device enforcement)Recalled 2025-09-05class ii

Spacelabs Healthcare, Inc.: Brand Name: Branding is not applied to this product. Product Name: DVI display cables Model/Catalog Number: 012-0895-12 Software Version: N/A Product Description: 12 foot D

Hazard

Due to display cables used for monitors were not in compliance with electromagnetic interference from inadequately shielded cables which may pose serious risk to patient's safety by interfering with the operation of another electronically sensitive medical device are in use.

Details from the source

Reason for recall: Due to display cables used for monitors were not in compliance with electromagnetic interference from inadequately shielded cables which may pose serious risk to patient's safety by interfering with the operation of another electronically sensitive medical device are in use. Distribution: Worldwide - United States Nationwide distribution including in the states of AK, AL, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, VA, WA, WI, WY and the countries of AFG, ARE, ARG, AUS, AUT, BGD, BHR, BOL, BRA, BRN, CAN, CHE, CHL, CHN, COL, CRI, CYP, CZE, DOM, DZA, ECU, EGY, ESP, FRA, GBR, GR Quantity: 820 Lot/code info: Lot Code: UDI-DI for products that are used with the display cables. (01)10841522131479 (Xprezzon), (01)10841522131455 (Qube)

UPC codes

10841522131479, 10841522131455

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.