There is a potential safety issue associated with the magnet venting paths of the Siemens 3 Tesla MRI system within your IMRIS Operating Suite. Siemens notified IMRIS that an ice blockage may exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed vent paths, leading to a pressure build-up within the helium containment system which could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
Reason for recall: There is a potential safety issue associated with the magnet venting paths of the Siemens 3 Tesla MRI system within your IMRIS Operating Suite. Siemens notified IMRIS that an ice blockage may exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed vent paths, leading to a pressure build-up within the helium containment system which could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room. Distribution: Worldwide distribution - US Nationwide and the countries of Canada, Sweden, Switzerland, Japan, Australia, Qatar, and China. Quantity: N/A Lot/code info: IMRIS Part Number(s): 105118-000, UDI/DI None predates UDI; 105118-001, UDI/DI None predates UDI; 105117-003, UDI/DI None predates UDI; 105117-004, UDI/DI None predates UDI; 105048-000, UDI/DI None predates UDI; 105048-001, UDI/DI None predates UDI; 105103-001, UDI/DI None predates UDI; 105071-004, UDI/DI None predates UDI; 105824-000, UDI/DI 00857534006189; 105828-000, UDI/DI 00857534006165; 116189-000, UDI/DI 00857534006202; 118396-000, UDI/DI 00857534006196.
00857534006189, 00857534006165, 00857534006202, 00857534006196
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.