Inadvertent energy delivery from surgical system instrument if 1) Force bipolar and bipolar instruments installed on system 2) Force bipolar connected to Force Triad generator, bipolar not connected to generator 3) Yellow pedal associated with force bipolar is pressed, released with head in surgeon console 4) Blue pedal associated with bipolar is pressed, resulting in force bipolar energy delivery
Reason for recall: Inadvertent energy delivery from surgical system instrument if 1) Force bipolar and bipolar instruments installed on system 2) Force bipolar connected to Force Triad generator, bipolar not connected to generator 3) Yellow pedal associated with force bipolar is pressed, released with head in surgeon console 4) Blue pedal associated with bipolar is pressed, resulting in force bipolar energy delivery Distribution: US: MS, PA, WI, KY, TN, MI, FL, GA, NC, NV, LA, OH, IL, MO, IN, CA, AK, MD, TX, IA, MA, NE, AZ, WA, CO, AL, SC Quantity: 162 Lot/code info: Software version P10 (OS4 v10.0.0/A70_P10_B738) SOFTWARE,EMBEDDED RLS,IS4000 AND IS4200,A70_P10_B738 (Model: 610092-738) on Model Name/Model Number/UDI-DI: ASSY,PSCART,IS4000,4-ARM,P10/380652-55/00886874110720, ASSY,SSC,IS4000,P10/380677-28/00886874110744, ASSY,VSS VISION SYSTEM,IS4000,P10/381121-45/00886874110898, ASSY,PSCART,IS4200,P10/380620-47/00886874115404, ASSY,VSC,IS4000,RECONDITIONED,P10/380721-15/00886874115343, ASSY,SSC,IS4000,RECONDITIONED,P10/380723-16/00886874115374
00886874110720, 00886874110744, 00886874110898, 00886874115404, 00886874115343, 00886874115374
Official record: FDA (Device enforcement) ↗