FDA (Device enforcement)Recalled 2024-08-21class ii

CHANGE HEALTHCARE CANADA COMPANY: Change Healthcare Cardiology Hemo. Versions 14.2, 14.1.1, 14.2, 14.3, 14.3.2, 14.3, 15.0.1, 15.1.

Hazard

A potential issue has been identified in Change Healthcare Cardiology Hemo where incorrect data entry can lead to inaccurate hemodynamic calculations, potentially resulting in misdiagnosis and inappropriate treatment. This issue occurs when the unit of measurement (UOM) for hemoglobin is configured in a way that differs from the users understanding.

Details from the source

Reason for recall: A potential issue has been identified in Change Healthcare Cardiology Hemo where incorrect data entry can lead to inaccurate hemodynamic calculations, potentially resulting in misdiagnosis and inappropriate treatment. This issue occurs when the unit of measurement (UOM) for hemoglobin is configured in a way that differs from the users understanding. Distribution: Worldwide distribution - US Nationwide and the countries of Australia, Canada, Germany, Ireland, Israel, United Kingdom. Quantity: 221 systems Lot/code info: UDI Numbers: (01)17540262100013(10)140100, (01)17540262100020(10)140101, (01)17540262100037(10)140200, (01)17540262100044(10)140300, (01)17540262100051(10)140302, (01)17540262100099(10)150000, (01)17540262100105(10)150001, (01)17540262100129(10)150100. Versions 14.2, 14.1.1, 14.2, 14.3, 14.3.2, 14.3, 15.0.1, 15.1.

UPC codes

17540262100013, 17540262100020, 17540262100037, 17540262100044, 17540262100051, 17540262100099, 17540262100105, 17540262100129

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify