An increase in clip locking malfunctions, Establish Final Arm Angle and Clip Opening While Locked, have been observed. Contributing causes may be clip locking component material properties and certain use conditions. Unsuccessful clip procedure can lead to recurrence of severe mitral regurgitation, device embolization, or tissue injury.
Reason for recall: An increase in clip locking malfunctions, Establish Final Arm Angle and Clip Opening While Locked, have been observed. Contributing causes may be clip locking component material properties and certain use conditions. Unsuccessful clip procedure can lead to recurrence of severe mitral regurgitation, device embolization, or tissue injury. Distribution: Worldwide distribution - US Nationwide including in the states of PR, WA, NE, MO, IA, KS, TN, MN, ND, MD, VA, MA, PA, DC, LA, KY, CA, NV, IL, WI, TX, CT, VT, AL, AR, GA, MS, MI, OH, FL, NC, NJ, CO, NY, DE, OR, SC, IN, HI, AZ, NH, ME, WV, OK, ID, UT, MT, NM, RI, AK, SD and the countries of AE, AR, AT, AU, BE, BG, BR, BY, CA, CH, CL, CN, CO, CR, CY, CZ, DE, DK, EC, EE, EG, ES, FI, FR, GB, GR, HK, HU Quantity: 20,045 Lot/code info: Part/UDI-GTIN: CDS0601-NTR/08717648226342, CDS0602-NTR/08717648226359, CDS0602-XTR/08717648226373, CDS0702-NT/08717648230950, CDS0702-NTW/08717648230974, CDS0702-XT/08717648230998, CDS0702-XTW/08717648231018, CDS0705-NT/08717648288616, CDS0705-NTW/08717648288630, CDS0705-XT/08717648288623, CDS0705-XTW/08717648288647, TCDS0202-XT/08717648229978, TCDS0202-NT/08717648229985, TCDS0205-XT/08717648288708, TCDS0205-NT/08717648288692, TCDS0302-NT/08717648334238, TCDS0302-NTW/08717648334252, TCDS0302-XT/08717648334245, TCDS0302-XTW/08717648334269, TCDS0305-NT/08717648334313, TCDS0305-NTW/08717648334337, TCDS0305-XT/08717648334320, TCDS0305-XTW/08717648334344, TCDS0202-NT/08717648229985, TCDS0205-NT/08717648288692 All lots.
08717648226342, 08717648226359, 08717648226373, 08717648230950, 08717648230974, 08717648230998, 08717648231018, 08717648288616, 08717648288630, 08717648288623, 08717648288647, 08717648229978, 08717648229985, 08717648288708, 08717648288692, 08717648334238, 08717648334252, 08717648334245, 08717648334269, 08717648334313, 08717648334337, 08717648334320, 08717648334344
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.