← Search recalls
FDA (Device enforcement)Recalled 2023-07-06class ii

Beckman Coulter, Inc.: Access Thyroglobulin Calibrators

Hazard

It has been determined through an internal investigation that the alternate Access Thyroglobulin calibrator cards for use with the Thyg reagent packs distributed in April 2023 will cause a decrease of approximately 11% in patient results.

Details from the source

Reason for recall: It has been determined through an internal investigation that the alternate Access Thyroglobulin calibrator cards for use with the Thyg reagent packs distributed in April 2023 will cause a decrease of approximately 11% in patient results. Distribution: worldwide except US, Canada, and China Quantity: 5980 units Lot/code info: UDI/DI 15099590227166, Serial/Lot Numbers: 233806, 234171, 234355, 234443, 234444, 234617, 337940

UPC codes

15099590227166

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.