← Search recalls
FDA (Device enforcement)Recalled 2023-08-18class ii

Alcon Research, LLC: Constellation Vision System, REF: 8065751150

Hazard

Ophthalmic microsurgical vision systems have a new Printed Circuit Board (PCB) design that could lead to higher voltage to a PCB component, that could cause component failure and loss of pressure control for the use of cutters, scissors, and forceps, which could lead to an increased risk of microbial infection, tissue damage and progressive visual impairment.

Details from the source

Reason for recall: Ophthalmic microsurgical vision systems have a new Printed Circuit Board (PCB) design that could lead to higher voltage to a PCB component, that could cause component failure and loss of pressure control for the use of cutters, scissors, and forceps, which could lead to an increased risk of microbial infection, tissue damage and progressive visual impairment. Distribution: US: GA, IL, PR Quantity: 3 Lot/code info: UDI-DI: 00380657511501, Serial Numbers: 1403028701X, 1202865201X, 0802983301X

UPC codes

00380657511501

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify