Beckman Coulter has become aware of an increase in customer complaints for intermittent IRISpec CB negative glucose control failure, resulting in false positive control results when used in conjunction with iChemVELOCITY Urine Chemistry Strips (PNs 800-7204 and 800-7212) on the iChemVELOCITY Analyzer. This issue may delay the reporting of patient results, with a worst-case scenario of a delay in diagnosis or treatment of metabolic disorders, kidney function abnormalities, urinary tract infections, and liver function.
Reason for recall: Beckman Coulter has become aware of an increase in customer complaints for intermittent IRISpec CB negative glucose control failure, resulting in false positive control results when used in conjunction with iChemVELOCITY Urine Chemistry Strips (PNs 800-7204 and 800-7212) on the iChemVELOCITY Analyzer. This issue may delay the reporting of patient results, with a worst-case scenario of a delay in diagnosis or treatment of metabolic disorders, kidney function abnormalities, urinary tract infections, and liver function. Distribution: worldwide Quantity: 5589 units Lot/code info: UDI/DI 10837461002611, Lot Numbers: 432-24 and higher
10837461002611
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.