FDA (Device enforcement)Recalled 2025-09-15class ii

FUJIFILM Healthcare Americas Corporation: FDR Visionary Suite, CH-200 Model/Version Numbers: (1) 566-16130-23, (2) 566-16130-31, (3) 566-16130-33

Hazard

It was found that the shaft for mounting the X-ray tube unit on the support may break. In addition, it was found that the protective parts might not work effectively only when the shaft was broken while being operated toward the front while lifting.

Details from the source

Reason for recall: It was found that the shaft for mounting the X-ray tube unit on the support may break. In addition, it was found that the protective parts might not work effectively only when the shaft was broken while being operated toward the front while lifting. Distribution: Domestic: AK, CA, FL, IL, MA, NH, OH, WI. Quantity: 16 units Lot/code info: Model/Version Number: (1) 566-16130-23, (2) 566-16130-31, (3) 566-16130-33; UDI-DI: (1) 04540217052226, (2) 04540217057436, (3) 04540217057450; Serial Numbers: (1) MP95A8F5A001, MP95A9F6A001, MP95A9F6B001, MP95A9F74001, MP95A9F74002, MP95A9F76001, MP95A9F78001, (2) MP95AA783001, (3) MP95AA983001, MP95AA983002, MP95AA983003, MP95AA985001, MP95AA985003, MP95AA985002, MP95AA988001, MP95AA988002;

UPC codes

04540217052226, 04540217057436, 04540217057450

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify