Luminex Molecular Diagnostics Inc: NxTAG Respiratory Pathogen Panel +SARS-CoV-2, REF: I056C0468

Hazard

Respiratory pathogen panel may report an elevated rate (7.6% predicted probability of occurrence) of non-specific hMPV target (human Metapneumovirus) results, risk identified predominantly in clinical samples that present positivity for another target, leading to an erroneous co-infection result, can potentially result in hMPV false positives.

Details from the source

Reason for recall: Respiratory pathogen panel may report an elevated rate (7.6% predicted probability of occurrence) of non-specific hMPV target (human Metapneumovirus) results, risk identified predominantly in clinical samples that present positivity for another target, leading to an erroneous co-infection result, can potentially result in hMPV false positives. Distribution: U Nationwide distribution in the states of OH, CA, NC, WI, TN, TX, MI, NY, MO, FL, MA, NJ, PA, VA, AZ, IL, OK, CO, MN. Quantity: 613 Lot/code info: Lot/Expiration: IK056C-0033/ 9/30/2024, IK056C-0034/ 10/31/2024, IK056C-0035/ 10/31/2024, IK056C-0036/ 12/31/2024, IK056C-0037/ 12/31/2024, IK056C-0038/ 12/31/2024, IK056C-0039/ 1/31/2025

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify