← Search recalls
FDA (Device enforcement)Recalled 2023-04-19class ii

Philips North America: Patient Information Center iX, Product Code 866386, and Patient Information Center iX Expand, Product Code 866390, Software Revision 4.0.1

Hazard

During extended operation of the MX40 with the PIC iX, the DHCP (Dynamic Host Configuration Protocol) lease can expire on the MX40. If the DHCP lease expires while the device is offline, upon reconnecting with the PIC iX, a new IP address will be assigned to the MX40. The PIC iX will not recognize the new IP address after reconnecting and therefore the expecting settings sync fails.

Details from the source

Reason for recall: During extended operation of the MX40 with the PIC iX, the DHCP (Dynamic Host Configuration Protocol) lease can expire on the MX40. If the DHCP lease expires while the device is offline, upon reconnecting with the PIC iX, a new IP address will be assigned to the MX40. The PIC iX will not recognize the new IP address after reconnecting and therefore the expecting settings sync fails. Distribution: Worldwide - US Nationwide distribution. Quantity: 591 devices Lot/code info: UDI-DI: 00884838104594; Serial Numbers:0Y5Z-7FHX-N 5H20-1DNN-E 1762-74NU-5 1C0V-08NR-0 6U53-30PD-1 6M5J-0JLM-H 2M2G-15LB-E 4U3A-3JGK-K 0439-16GM-G 3A1T-22HW-U 534U-57JE-Y 0D6U-5ZH1-D 3H5C-6JK9-X 3K0C-3EH3-H 3Z47-12JD-V 524N-43JR-2 5N40-2RHR-N 1331-5BNT-B 1V4F-34MB-G 1B4R-37N1-3 6D3L-28MC-M 7Y52-29H4-L 7Y5Z-7FHX-F 1X6W-21KC-W 6278-09GU-Y 2C6N-3KLU-C 3E5Y-32PN-4 5T2L-2TP9-V 021Z-1WJ2-5 1W6W-21KC-Z 7L3H-46MR-L 194D-2YHB-T 1Y7A-49M4-6 2Z3C-1AHM-3 4T51-53NK-A 1P4A-44NB-V 3B5M-55J9-5 5V57-0HN3-J 7V5Z-7FHX-R 224B-2HJF-F 5G1G-01JC-Z 6Y7F-06MT-0 676T-6XMZ-M 5F1G-01JC-2 5K2B-6DHP-4 5T45-7ELE-C 0V7B-4EJH-A 1F23-36KM-0 2C50-6KKD-R 3107-4RG3-6 3X5B-1NH8-2 6U32-24GX-0 5K2L-2TP9-D 4T2R-6HHJ-D 613L-28MC-T 7K21-0XNW-C 062U-1VNP-B 5M78-09GU-6 6226-14LG-X 0V6R-0PLF-7 5K72-2NL7-J 7G21-0XNW-M 077M-7JG6-L 2R4Y-5JHY-8 7Z4E-0EG2-N 1X2J-6LPP-R 1Y0A-64MU-A 0Z3E-41PZ-K 327Y-2CJW-6 3979-6CPV-0 2150-6KKD-T 4E6Z-70HU-B 1H3G-4UJL-D 2A75-27LY-3 3X3J-5PP3-5 4V7G-0BLD-8 567W-7BG9-5 5K6K-0RGF-L 6F2V-1BLC-W 0106-0UH7-D 1T1M-1TLH-F 5N2E-2FMY-1 057M-7JG6-U 0M5T-3FHA-Y 4B7K-7MPL-Y 4J2R-6HHJ-2 6H7F-06MT-7 184A-44NB-8 2518-4CLX-9 4Y40-2RHR-Z 5Y6J-3HL1-E 241K-18LJ-K 3R04-4ZPR-B 5B5G-0ZNF-T 4N77-1HM0-D 7W49-4XM7-M 723D-20JM-C 2V2W-3VM1-X 4T12-4BPG-8 5V3L-28MC-C 6J2K-47P7-C 2975-27LY-6 2E48-7GN2-3 2K0X-0TLA-F 302F-7YK1-P 3215-4JK4-E 3B05-6RJX-W 530G-60JV-Z 5843-5RM5-X 5J6K-0RGF-P 5U6T-6XMZ-X 684L-2BNM-Y 6K5U-66H9-G 743H-46MR-4 783K-6VGP-M 7F1U-7TJA-7 7U09-3WPK-T 0Z69-7VM6-F 1063-19MD-6 114M-03P1-2 1Y1E-50P2-J 2P5C-6JK9

UPC codes

00884838104594

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.