Access hsTnI High Sensitivity Troponin I, REF B52699 Access Immunoassay System REAGENT PACK
Check against the official notice below — it lists the full affected range.
Beckman Coulter Inc. is recalling 293,640 kits of its Access hsTnI High Sensitivity Troponin I reagent pack (catalog number B52699) distributed across the United States, U.S. territories, and numerous international locations. The recall was initiated based on customer reports indicating that samples with very high cardiac troponin I concentrations exceeding 270,000 pg/mL can cause contamination of the reagent pack through sample carryover. This contamination occurs when the testing probe transfers sample material into the particle well of the reagent pack, with the extent of carryover depending on both the sample concentration level and how frequently high concentration samples are tested.
The affected product includes multiple lot numbers, with specific UDI codes and expiration dates listed in the recall documentation. According to the recall record, no remedy information is specified for consumers who may have affected kits.
Written by software from the official notice below. Not reviewed by a lawyer.
Through customer feedback, testing a sample with cardiac troponin I concentration that are >270,000 pg/mL may cause carryover into the hsTnl reagent pack. The carryover is caused by sample-to-pack carryover predominantly into the particle well of the reagent pack by way of the probe. The particle well then becomes contaminated by the high sample from the probe. The extent of the carryover is dependent on sample concentration and frequency of high concentration samples.
Reason for recall: Through customer feedback, testing a sample with cardiac troponin I concentration that are >270,000 pg/mL may cause carryover into the hsTnl reagent pack. The carryover is caused by sample-to-pack carryover predominantly into the particle well of the reagent pack by way of the probe. The particle well then becomes contaminated by the high sample from the probe. The extent of the carryover is dependent on sample concentration and frequency of high concentration samples. Distribution: U.S.: AL, AR, AZ, CA, CO, CT, DE, DC, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, ME, MD, MI, MN, MS, MO, NC, ND, NE, NV, NJ, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, VA, WA, WI, WV, and WY O.U.S.: Algeria, Andorra, Australia, Austria, Bahrain, Bangladesh, Belgium, Bosnia and Herzegovina, Brazil, Brunei, Bulgaria, Canada, China, Croatia, Cura¿ao, Czech Republic, Denmark, Egypt, Eston Quantity: 293,640 kits Lot/code info: Catalog Number:B52699 Lot Number / UDI Code: 921491 / (01)15099590693183(17)201130(11)191201(10)921491; 921638 / (01)15099590693183(17)201231(11)200101(10)921638; 921631 / (01)15099590693183(17)210131(11)200201(10)921631; 921750 / (01)15099590693183(17)210131(11)200201(10)921750; 921805 / (01)15099590693183(17)210228(11)200229(10)921805; 921831 / (01)15099590693183(17)210331(11)200331(10)921831; 921943 / (01)15099590693183(17)210430(11)200430(10)921943; 922000 / (01)15099590693183(17)210531(11)200531(10)922000; 922039 / (01)15099590693183(17)210630(11)200630(10)922039; 922040 / (01)15099590693183(17)210731(11)200731(10)922040; 922062 / (01)15099590693183(17)210731(11)200731(10)922062; 922145 / (01)15099590693183(17)210831(11)200831(10)922145; 922146 / (01)15099590693183(17)210930(11)200930(10)922146; 922338 / (01)15099590693183(17)210930(11)200930(10)922338; 922440 / (01)15099590693183(17)211031(11)201031(10)922440; 922545 / (01)15099590693183(17)211031(11)201031(10)922545; 922547 / (01)15099590693183(17)211130(11)201130(10)922547; 922713 / (01)15099590693183(17)211231(11)201231(10)922713; 922714 / (01)15099590693183(17)211231(11)201231(10)922714; 922722 / (01)15099590693183(17)220131(11)210131(10)922722; 922897 / (01)15099590693183(17)220131(11)210131(10)922897; 124235 / (01)15099590693183(17)220228(11)210228(10)124235; 124339 / (01)15099590693183(17)220228(11)210228(10)124339; 922898 / (01)15099590693183(17)220228(11)210228(10)922898; 124340 / (01)15099590693183(17)220331(1