← Search recalls
FDA (Device enforcement)Recalled 2024-08-01class ii

Outset Medical, Inc.: Tablo Console, REF: PN-0003000, PN-0006000, part of the Tablo Hemodialysis System. The Tablo Hemodialysis System is indicated for use in patients with acute and/or chronic renal

Hazard

Blood leak sensor, a component of hemodialysis device, may show decreased sensitivity, at the lowest level, specified as 0.35 mL/min blood leak rate, which could delay triggering the ALARM_DIALYZER_BLOOD_LEAK alarm, if blood loss, less-than 300ml, occurs, the patient may experience hypotension, fatigue, cramps, dizziness, headaches, nausea, and shortness of breath.

Details from the source

Reason for recall: Blood leak sensor, a component of hemodialysis device, may show decreased sensitivity, at the lowest level, specified as 0.35 mL/min blood leak rate, which could delay triggering the ALARM_DIALYZER_BLOOD_LEAK alarm, if blood loss, less-than 300ml, occurs, the patient may experience hypotension, fatigue, cramps, dizziness, headaches, nausea, and shortness of breath. Distribution: US: CA, NY, IL, MO, KY, AZ, PA, FL, GA, AK, NM, VA, NC, SC, OK, AL, MA, OH, MI, WA, ID, TX, TN, WY, IA, LA, NE, IN, AR, NV, WV, CO, NJ, MD, KS, UT, CT, MN, MS, WI, HI, DE, VT, MT, OR, ND. OUS: UAE Quantity: 2,422 Lot/code info: UDI-DI: 00850001011112, 00850001011204. Serial Numbers: 400405, 602382M, 602384M, 602387M, 602400M, 602402M, 601124M, 601156M, 601102M, 601265M, 400109, 602248M, 602659M, 602670M, 602767M, 602769M, 602771M, 601356M, 602421M, 602403M, 400394, 601728M, 601730M, 601731M, 601733M, 601735M, 601736M, 601738M, 601771M, 601772M, 601774M, 600599M, 600614M, 600615M, 600620M, 600631M, 600641M, 600646M, 600648M, 600654M, 600659M, 601337M, 601357M, 601360M, 601361M, 601364M, 601368M, 601384M, 601395M, 601396M, 601422M, 601727M, 602748M, 600189, 601515M, 601513M, 601514M, 601517M, 601518M, 601838M, 601864M, 601870M, 601915M, 601924M, 601823M, 601854M, 601861M, 601887M, 601901M, 601908M, 601923M, 601903M, 601936M, 601945M, 601974M, 601799M, 600484M, 601876M, 601886M, 601938M, 601941M, 600488M, 601129M, 601146M, 601270M, 601298M, 601797M, 601805M, 601812M, 601814M, 601821M, 601855M, 601890M, 601897M, 601932M, 601934M, 600563M, 600569M, 600572M, 600574M, 601952M, 601954M, 601878M, 601899M, 601935M, 601942M, 601943M, 601947M, 601800M, 601811M, 601848M, 601874M, 601516M, 602354M, 600174M, 601955M, 602025M, 602035M, 601598M, 601620M, 601139M, 601149M, 601209M, 601210M, 601218M, 601842M, 602503M, 600261M, 602570M, 602626M, 602629M, 601857M, 601858M, 601862M, 601863M, 601866M, 601709M, 602185M, 602255M, 602257M, 602675M, 602744M, 602747M, 602750M, 602753M, 602754M, 602755M, 602756M, 602757M, 602759M, 601639M, 601640M, 601641M, 601642M, 601643M, 601648M, 602053M, 602075M, 602735M, 602784M, 602798M,

UPC codes

00850001011112, 00850001011204

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.