← Search recalls
FDA (Device enforcement)Recalled 2025-09-03class ii

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.: Allura Xper FD20/15; Model Numbers: 722058; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;

Hazard

The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.

Details from the source

Reason for recall: The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted. Distribution: Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Curacao, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Guinea, H Quantity: 300 units (69 US, 231 OUS) Lot/code info: Model Numbers: 722058; UDI-DI: 00884838059115; Serial Numbers: 337, 178, 249, 22, 202, 279, 254, 285, 71, 186, 82, 85, 162, 138, 24, 365, 203, 167, 146, 168, 304, 172, 200, 123, 129, 149, 259, 4, 197, 134, 38, 290, 171, 56, 264, 238, 199, 61, 156, 99, 65, 59, 109, 100, 305, 319, 215, 278, 124, 196, 405, 325, 265, 378, 277, 274, 94, 182, 366, 340, 190, 219, 373, 239, 23, 205, 257, 212, 111, 110, 399, 72, 332, 92, 142, 67, 106, 335, 360, 415, 11, 114, 151, 93, 232, 395, 292, 380, 183, 349, 261, 15, 375, 161, 78, 368, 9, 372, 288, 252, 320, 213, 293, 389, 169, 97, 336, 25, 143, 133, 192, 49, 412, 201, 89, 66, 163, 170, 77, 273, 35, 222, 229, 248, 334, 384, 40, 268, 240, 258, 289, 269, 226, 244, 48, 141, 391, 241, 164, 381, 281, 154, 246, 127, 315, 256, 233, 348, 204, 396, 7, 185, 95, 131, 377, 60, 68, 206, 101, 150, 333, 84, 398, 313, 3, 39, 344, 376, 382, 370, 387, 419, 394, 407, 413, 392, 411, 358, 409, 385, 416, 367, 379, 331, 418, 356, 421, 369, 420, 397, 357, 408, 393, 388, 108, 211, 126, 125, 301, 12, 30, 355, 354, 159, 139, 128, 81, 58, 96, 343, 177, 115, 19, 18, 311, 220, 386, 291, 327, 214, 145, 223, 137, 136, 119, 103, 404, 27, 32, 350, 400, 422, 423, 347, 73, 43, 55, 194, 189, 237, 230, 221, 247, 116, 242, 29, 37, 276, 345, 166, 179, 53, 87, 303, 90, 112, 245, 148, 41, 234, 208, 117, 193, 76, 236, 266, 130, 235, 251, 86, 31, 195, 144, 330, 63, 295, 70, 187, 79, 10, 33, 75, 361, 153, 284, 102, 21, 64, 45, 36, 181, 351, 272, 210, 243, 20, 98, 34, 16, 218;

UPC codes

00884838059115

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.