← Search recalls
FDA (Device enforcement)Recalled 2022-08-25class ii

Zimmer, Inc.: G7 Acetabular System, Acetabular Shell, Item numbers Item Numbers: 010000663 (52 MM E), 010000667 (60 MM G). used in hip arthroplasty

Hazard

The outer sterile package cavity has a corner wall thickness that is below the specification. The thin corner wall could potentially crack during transit. Potential risks include non-clinically or clinically significant extension of surgery, or infection leading to surgical intervention.

Details from the source

Reason for recall: The outer sterile package cavity has a corner wall thickness that is below the specification. The thin corner wall could potentially crack during transit. Potential risks include non-clinically or clinically significant extension of surgery, or infection leading to surgical intervention. Distribution: Domestic distribution US nationwide. Quantity: 56 US Lot/code info: Item No. 010000663 UDI-DI: 00880304524224 Lots: 7286477 7289929 7296576 7296633 7301697 Exp. 21-Jul-32 Item No. 010000667 UDI-DI: 00880304524262 Lot: 7253325 Exp. 13-Jun-32

UPC codes

00880304524224, 00880304524262

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.